SHANDONG SCIENCE ›› 2015, Vol. 28 ›› Issue (1): 43-46.doi: 10.3976/j.issn.1002-4026.2015.01.007

• Biosensors • Previous Articles     Next Articles

Analdehyse oxidase modified gold electrode based formaldehyde gas sensor

GAO Guangheng,BI Chunyuan,ZHU Sirong, ZHOU Wanli,   

  1. 1. Shandong Provincial Key Laboratory of Biosensors, Biotechnology Center, Shandong Academy of Sciences, Jinan 250014, China; 2. Institute of Biology, Shandong Academy of Sciences, Jinan 250014, China
  • Received:2014-05-26 Online:2015-02-20 Published:2015-02-20

Abstract: We designed a formaldehyde gas detection sensor based on the aldehyde oxidase principle. Its sensitive component was manufactured through such process as selfassembly, activation, and adsorption with gold electrode as substrate, and had better electrocatalysis and oxidation effect to formaldehyde. We further designed a formaldehyde gas detection biosensor with the electrode and a LPC1788 chip as its CPU. The biosensor was tested with formaldehyde gas of 0.1 mg/m3.Results demonstrate that the sensor has as rapid response time as 20 s, shorter recovery time and lower detection limit as 0.01 mg/m3 to formaldehyde gas.

Key words: aldehyde oxidase, gold electrode, biosensor, LPC1788

CLC Number: 

  • TP274