SHANDONG SCIENCE ›› 2014, Vol. 27 ›› Issue (4): 8-11.doi: 10.3976/j.issn.1002-4026.2014.04.002

• Oceanographic Science, Technology and Equipment • Previous Articles     Next Articles

Design of highaccuracy temperature measurement circuit of an expendable bathythermograph

 LEI Zhuo, HE Hai-Jing, QU Jun-Le, LIU Jie   

  1. Shandong Provincial Key Laboratory of Ocean Environmental Monitoring Technology, Institute of Oceanographic
     Instrumentation, Shandong Academy of Sciences, Qingdao 266001, China
  • Received:2014-06-03 Online:2014-08-20 Published:2014-08-20

Abstract:

This paper devises a highaccuracy temperature measurement circuit with a temperature measurement bridge to improve temperature measurement accuracy of an expendable bathythermograph (XBT). A new calibration method is also presented. We further design a reliable water immersion detection module combined with software and hardware. Ocean experiment results indicate that the temperature curve measured by XBT has more than ninety percent similarity degree with standard CTD, less than 0.2 ℃ for average temperature measurement error. This is in the range of allowed error, so temperature measurement accuracy attains expected values.

Key words: expendable bathythermograph, temperature measurement accuracy, water immersion detection

CLC Number: 

  • P715.5