Shandong Science ›› 2020, Vol. 33 ›› Issue (6): 118-125.doi: 10.3976/j.issn.1002-4026.2020.06.016

• Other Research Article • Previous Articles     Next Articles

Design of dual-channel non-dispersive infrared SO2online detection system

LIU Zhi-feng, LUO Xue-ke, LI Wen, LIU Peng   

  1. School of Mechanical and Materials Engineering, North China University of Technology, Beijing 100144
  • Received:2020-03-30 Online:2020-12-09 Published:2020-12-11

Abstract: The current problems of multiple interference factors and slow response time in online detection of low-volume-fraction SO2 gas in flue gas detection.In this case,an online SO2 detection system based on the non-dispersive infrared absorption principle was designed,which used the characteristic capability of SO2 to absorb infrared light in a specific band and adopted a single light source and dual channel method. Experiment results show that the zero drift is 0.4%, the response time is 3.86 s, the linear error is 0.3%, and the repeatability error is 0.52%. A comparison of parameters shows that this method can accurately measure SO2 concentration and is more suitable for the current demand of online flue gas detection.

Key words: SO2, nondispersive infrared absorption, single light source and dual channel, response time

CLC Number: 

  • O657.32